FDA Recall Terminated

MPI Rapid Response HUT Table, Model #RRHUT 1017. The firm name on the label is Medical Positioning Inc., Kansas City, MO. Designed for multiple uses including a conventional examination table, unexplained syncope, neurocardiogenic syncope, chronic fatigue syndrome, and most noninvasive procedures.

Recall: Z-1211-2012 · Initiated March 5, 2012

Recall

Recall Number
Z-1211-2012
Event Number
61348
Firm
Medical Positioning Inc
FEI Number
1000118382
Product Code
LGX
Status
Terminated
Root Cause
Error in labeling
Initiated
March 5, 2012
Posted
March 15, 2012
Terminated
July 11, 2013
Address
1717 Washington St, Kansas City, MO, 64108-1121

Description

MPI Rapid Response HUT Table, Model #RRHUT 1017. The firm name on the label is Medical Positioning Inc., Kansas City, MO. Designed for multiple uses including a conventional examination table, unexplained syncope, neurocardiogenic syncope, chronic fatigue syndrome, and most noninvasive procedures.

Reason

The warning label regarding electrical shock was not placed on the device warning to have the power switch in the "OFF" position prior to connecting the unit to electric power.

Action

Medical Positioning Inc. sent a "RAPID RESPONSE HUT TABLE LABELING CORRECTION NOTIFICATION" letter dated March 5, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter included an Owner's Manual insert page which clearly states the warning and how to avoid the electric shock hazard and a warning tag to attach to the product's power cord near the plug end using the provided wire tie. Contact the firm at 816-595-2356 for questions regarding this notice.

Distribution

Worldwide Distribution-USA (nationwide) including the states of IN, KS, WI, NY, IL, NC, OH, and MS and the country of Thailand.

Quantity

5 tables