FDA Recall
Terminated
XiO Radiation Treatment Planning System, Release 4.51.00.
Recall: Z-1209-2010
·
Initiated February 22, 2010
Recall
- Recall Number
- Z-1209-2010
- Event Number
- 55008
- Firm
- Computerized Medical Systems Inc
- FEI Number
- 1937649
- Product Code
- MUJ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- February 22, 2010
- Posted
- March 25, 2010
- Terminated
- August 24, 2011
- Address
- 13723 Riverport Drive, Suite 100, Maryland Heights, MO, 63043
Description
XiO Radiation Treatment Planning System, Release 4.51.00.
Reason
Inconsistent MLC positions, dose, and monitor units
Action
Elekta contacted customers by phone and mail beginning February 22, 2010, explaining the affected product and providing software update information. For further information, contact Elekta at 1-408-830-8023.
Distribution
Worldwide Distribution -- United States (PA, AZ, CA, MN, OH, WI, NC, FL, TX, NY and WI), Australia, The Netherlands and India.
Quantity
Planning system is loaded 25 in locations