Rechargeable Li-ion battery used with Puritan Bennett TM 980 Series Ventilator System including Pediatric-Adult Ventilator, Neonatal Ventilator, Universal Ventilator; Battery Product Number: 10086042; UDI: 20884521700861 The rechargeable lithium-ion batteries are intended to provide a backup power source to the ventilator in case AC power is lost.
Recall
- Recall Number
- Z-1208-2018
- Event Number
- 79371
- Firm
- Covidien LLC
- FEI Number
- 1219930
- Product Code
- CBK
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- November 21, 2017
- Terminated
- July 10, 2020
- Address
- 60 Middletown Ave, North Haven, CT, 06473-3908
Description
Rechargeable Li-ion battery used with Puritan Bennett TM 980 Series Ventilator System including Pediatric-Adult Ventilator, Neonatal Ventilator, Universal Ventilator; Battery Product Number: 10086042; UDI: 20884521700861 The rechargeable lithium-ion batteries are intended to provide a backup power source to the ventilator in case AC power is lost.
Rechargeable lithium-ion batteries with incorrect firmware that are used in certain Puritan Bennett 980 ventilators may not fully charge after installation.
Consignees are instructed to immediately notify all care environments in which the PB980 ventilators are used about the notification. If the facility has distributed PB980 ventilators to other persons or facilities, they are asked to please promptly forward a copy of the letter to those recipients. The consignee is asked to complete an attached form and return it as directed to confirm receipt and understanding of the notification.The firm developed a tool to inspect and identify the firmware on the batteries. Medtronic service engineers will inspect batteries at consignee facilities during scheduled or preventive maintenance. Batteries found to have the incorrect firmware will be replaced. For further questions, please call (203) 492-5415.
Worldwide Distribution - USA (nationwide) Distribution to the states of : AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY, and the District of Columbia. and to the countries of : Argentina, Chile, Mexico, Panama, Canada, Australia, New Zealand, Austria, Bahrain, Belgium, Bulgaria, Croatia, Czech Republic, Egypt, Estonia, France, Germany, Greece, Hungary, Ireland, Italy, Jordan, Kuwait, Latvia, Lebanon, Lithuania, Morocco, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Israel, Netherlands, Tunisia, Brunei, Indonesia, Malaysia, Myanmar, Philippines, Singapore, Thailand, and Vietnam
15,885