FDA Recall Terminated

Screwdriver for 3.5mm Screws, DARCO system, instrument for use in orthopaedic implant surgery. REF: DC4261, Non-sterile, Rx only, 1 each, Wright Medical Technology, Inc., 5677 Airline, Road, Arlington, TN 38002

Recall: Z-1208-2009 · Initiated February 19, 2009

Recall

Recall Number
Z-1208-2009
Event Number
51431
Firm
Wright Medical Technology Inc
FEI Number
1043534
Product Code
LXH
Status
Terminated
Root Cause
Process design
Initiated
February 19, 2009
Posted
April 28, 2009
Terminated
February 3, 2010
Address
5677 Airline Rd, Arlington, TN, 38002

Description

Screwdriver for 3.5mm Screws, DARCO system, instrument for use in orthopaedic implant surgery. REF: DC4261, Non-sterile, Rx only, 1 each, Wright Medical Technology, Inc., 5677 Airline, Road, Arlington, TN 38002

Reason

The package insert lists 3 sterilization procedures, 2 of which are not validated. Pre-vacuum steam sterilization is validated.

Action

Wright Medical notified their distributors of the recall on 03/11/2009 by letter explaining the problem and requesting return of the product from their inventory and from their consignees. The letter included a "fax back" response page to ensure the receipt of the notice. The hospitals and surgeons were notified of the recall also on 03/11/2009. All letters were delivered by Fed Ex.

Distribution

Worldwide distribution: USA, Canada, Germany, France, Belgium, the Netherlands, Italy, Finland, Austria, Denmark, Norway, Turkey, South Africa, United Kingdom, and Australia.

Quantity

682 devices