Zoll AED Plus Defibrillator
Recall
- Recall Number
- Z-1206-2009
- Event Number
- 51053
- Firm
- ZOLL Medical Corporation, World Wide Headquarters
- FEI Number
- 1220908
- Product Code
- MKJ
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- February 12, 2009
- Posted
- April 1, 2009
- Terminated
- April 9, 2026
- Address
- 269 Mill Rd, Chelmsford, MA, 01824
Description
Zoll AED Plus Defibrillator
Device fails to discharge the defibrillation energy.
On 3/31/09, the firm revised their recall strategy and required all users to a perform a "Mandatory Upgrade" software The software upgrade allows all potentially affected devices to monitor battery charging performance through periodic self-testing. If defective batteries are detected at any time prior to the recommended maximum of five years, users are prompted by the device to install fresh batteries. Frim issued Press Release on April 2, 2009. In the first notification Zoll notified customers by letter on 2/12/09 via certified mail. Customers instructed to replace device batteries every three years. In lieu of replacing batteries every three years, customers will also have the choice to update their devices with software that will monitor batteries for the identified defect. ZOLL will be send email notification to all consignees who have provided an email address as part of their contact information. ZOLL will also be publishing information of the notification on industry magazines. http://www.zollaedplusbatteryhelp.com/
Worldwide distribution
183,535 units