FDA Recall Open, Classified

Sit to Stand STS500 Patient Lift

Recall: Z-1205-2023 · Initiated February 3, 2023

Recall

Recall Number
Z-1205-2023
Event Number
91695
Firm
Med-Mizer, Inc.
FEI Number
3004976058
Product Code
FSA
Status
Open, Classified
Root Cause
Device Design
Initiated
February 3, 2023
Posted
March 2, 2023
Address
80 Commerce Dr, Batesville, IN, 47006-6700

Description

Sit to Stand STS500 Patient Lift

Reason

Risk of boom pivot failing due to the nut securing the boom becoming loose and the bolt to be unsecure to the flange on the side.

Action

Med-Mizer notified distributors of this recall event by distributing an Urgent Medical Device Recall Notice on 02/03/2022 via Certified Mail. The MR600 and STS500 Patient Lifts are being subject to recall due to the potential for the nut securing the boom to become loose over time, which could result in the boom moving unexpectedly side to side or dropping the patient during operation. Customers are asked to inspect all affected devices at their facility to determine if the nut securing the boom is loose. If it is, immediately remove the device from service. Med-Mizer is providing customers with affected devices replacement parts in order to complete the correction. Customers are to complete the provided Correction Response Form to Med-Mizer and return via fax at (812)932-5678, by email at [email protected], or by mail at 80 Commerce Drive, Batesville, IN 47006. If your firm is a distributor, please provide this notice to your customers. Any questions can be directed to Med-Mizer by phone at (877)867-7365.

Distribution

Awaiting consignee list from the recalling firm.

Quantity

37 units