FDA Recall Terminated

PLAC Test Calibrator Kit, Five-level liquid calibrator set for use with the diaDexus PLAC Test Reagent Kit, assembled in a kit box with an outer label, Catalog numbers 90108 and 10-0108, manufactured by diaDexus inc., South San Francisco, CA. Intended to establish points of reference that are used in the determination of values in the determination of Lp-PLA2 by the PLAC Test Reagent Kit.

Recall: Z-1205-2009 · Initiated February 9, 2009

Recall

Recall Number
Z-1205-2009
Event Number
51215
Firm
DiaDexus, Inc
FEI Number
3003643666
Product Code
NOE
Status
Terminated
Root Cause
Process control
Initiated
February 9, 2009
Posted
April 23, 2009
Terminated
January 5, 2010
Address
343 Oyster Point Blvd, South San Francisco, CA, 94080

Description

PLAC Test Calibrator Kit, Five-level liquid calibrator set for use with the diaDexus PLAC Test Reagent Kit, assembled in a kit box with an outer label, Catalog numbers 90108 and 10-0108, manufactured by diaDexus inc., South San Francisco, CA. Intended to establish points of reference that are used in the determination of values in the determination of Lp-PLA2 by the PLAC Test Reagent Kit.

Reason

Premature Expiration-- the product is projected to fall short of expected shelf life, and may result in inaccurate results.

Action

Distributors and Direct accounts were notified by phone followed by Notification letter and certificate of destruction via FedEx. International customers were notified via e-mail with PDF copies of notices, to be followed by hard copies via FedEx. Customers are requested to record the number of kits (full and partial ) they had on the Certificate of Destruction and to dispose of that number of kits with two people signing the form, and then to fax that form to diaDexus.

Distribution

Worldwide Distribution

Quantity

105 kits