da Vinci S. Surgical System, da Vinci Si and da Vinci Si-e Surgical Systems. Intuitive Surgical, Inc. Sunnyvale, CA 94086. The Intuitive Surgical Endoscopic instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.

Recall: Z-1202-2012 · Initiated February 6, 2012

Recall

Recall Number
Z-1202-2012
Event Number
61156
Firm
Intuitive Surgical, Inc.
FEI Number
3001675293
Product Code
NAY
Status
Terminated
Root Cause
Device Design
Initiated
February 6, 2012
Posted
March 13, 2012
Terminated
May 9, 2012
Address
1266 Kifer Rd, Bldg 100, Sunnyvale, CA, 94086-5304

Description

da Vinci S. Surgical System, da Vinci Si and da Vinci Si-e Surgical Systems. Intuitive Surgical, Inc. Sunnyvale, CA 94086. The Intuitive Surgical Endoscopic instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.

Reason

The holding brake, a component of the da Vinci S Surgical System, may allow passive uncontrolled motion due to gravity during specific power-off conditions.

Action

Intuitive Surgical sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated February 3, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter provided new instructions for use. Additionally, a Device Correction Response Form was enclosed for customers to complete and return via fax to 408-523-1390. Contact Intuitive Surgical Customer Service at 1-800-876-1310 for questions regarding this notice.

Distribution

Worldwide Distribution.

Quantity

1966 systems are in commerce at hospital locations.