Spacelabs Medical 1400 MHz Telemetry System,. Model 91341-09. Telemetry System consists of Model 91341-09 transmitter, 90478-1CST receiver and 90478-1CSV receiver.

Recall: Z-1202-06 · Initiated May 5, 2006

Recall

Recall Number
Z-1202-06
Event Number
35391
Firm
Spacelabs Medical Incorporated
FEI Number
3010157426
Product Code
DSI
Status
Terminated
Root Cause
Other
Initiated
May 5, 2006
Posted
July 6, 2006
Terminated
December 27, 2010
Address
5150 220th Ave Se, Issaquah, WA, 98029-6834

Description

Spacelabs Medical 1400 MHz Telemetry System,. Model 91341-09. Telemetry System consists of Model 91341-09 transmitter, 90478-1CST receiver and 90478-1CSV receiver.

Reason

Potential for patient waveforms to move to an open receiver module or to an occupied receiver causing the intermittent inappropriate monitoring of both patients for several minutes.

Action

On May 5, 2006 a letter was issued to all consignees, via return receipt requested. The letter advised consignees of the issue and provided some ways to mitigate the risk. Anyone experiencing the anomaly is to document the failure and report it to the firm. A field service representative will visit each consignee to optimize the antenna system to reduce intermodulations. After determining the root cause and validating a fix, the firm issued a second letter dated August 21, 2006 to their consignees. The letter advised consignees of three issues that needed 'improvement' and stated a service engieneer will schedule a time to visit and install the improvements.

Distribution

The systems were distributed to 22 hospitals/medical centers nationwide and to 2 Spacelabs personnel. There is no international distribution.

Quantity

22 telemetry systems; 1229 transmitters