Hercules 360 Universal Stabilizing Arm, top mount, 360-degree rotation, reusable stainless steel
Recall
- Recall Number
- Z-1201-2017
- Event Number
- 76344
- Firm
- Terumo Cardiovascular Systems Corporation
- FEI Number
- 3000204839
- Product Code
- MWS
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- August 20, 2014
- Posted
- February 14, 2017
- Terminated
- February 15, 2017
- Address
- 125 Blue Ball Rd, Elkton, MD, 21921-5315
Description
Hercules 360 Universal Stabilizing Arm, top mount, 360-degree rotation, reusable stainless steel
Terumo Cardiovascular Systems (Terumo CVS) is recalling certain lots of Hercules TM 360 Universal Stabilizing Arms because the two pins that hold the sleeve attached to the distal end of the device may come loose, causing the pins and sleeve to separate from the device. While the arm can still function without the pins and sleeve in place, there is potential for the loose components to fall into the surgical site.
Terumo Cardiovascular Systems has contact all affected parties (user facilities and consignees) via a Customer Notification Letter that advises of the recall activity.
Terumo Cardiovascular has identified 16 domestic consignees and 30 international consignees .
The total number of distributed devices stands at 260