FDA Recall Terminated

Mar Cor Purification CWP Reverse Osmosis System. Models 101, 102, 104, and 106.

Recall: Z-1201-2014 · Initiated June 27, 2012

Recall

Recall Number
Z-1201-2014
Event Number
67516
Firm
Mar Cor Purification
FEI Number
3006475063
Product Code
FIP
Status
Terminated
Root Cause
Component design/selection
Initiated
June 27, 2012
Posted
March 13, 2014
Terminated
August 22, 2014
Address
14550 28th Ave N, Minneapolis, MN, 55447-4817

Description

Mar Cor Purification CWP Reverse Osmosis System. Models 101, 102, 104, and 106.

Reason

There exists a potential situation with the crimp connection at the pump head assembly. The potential exists that the reverse osmosis pump head wiring could build resistance resulting in resistive heating of the surrounding components. If this occurs, the pump head assembly may experience melting of the plastic components and potentially create smoke/fire and shut down of the unit.

Action

Mar Cor Technicians started visiting the consignees to make the correction, in June 2012 Mar Cor called the consignee prior to the visit. An Urgent Medical Device correction letter was sent to customers beginning 2/27/2014. the letter identified the affected units, described the issue, stated that a correction has been, or will be made to the devices and asked for a response form be sent back to the firm.

Distribution

nationwide. No distribution outside the US.

Quantity

130