FDA Recall Terminated

Abbott CELL-DYN 3500 Hematology Analyzer

Recall: Z-1199-2007 · Initiated May 11, 2007

Recall

Recall Number
Z-1199-2007
Event Number
38322
Firm
Abbott Laboratories
FEI Number
2919069
Product Code
GKZ
Status
Terminated
Root Cause
Other
Initiated
May 11, 2007
Posted
August 29, 2007
Terminated
June 19, 2008
Address
5440 Patrick Henry Dr, Santa Clara, CA, 95054-1113

Description

Abbott CELL-DYN 3500 Hematology Analyzer

Reason

The product failed electrostatic discharge testing, which could result in product failure.

Action

Mandatory Technical Service Bulletins (TSB) are being issued to Field Service to upgrade all instruments in the field. All customers will be contacted by Field Service to schedule this mandatory upgrade. The upgrades are expected to take 12 months.

Distribution

Worldwide, including USA, Puerto Rico, Canada, Mexico, Argentina, Australia, Bahamas, Brazil, Chile, China, Colombia, Costa Rica, Dominican Republic, El Salvador, Ecuador, Germany, Guatemala, Hong Kong, Jamaica, Japan, New Zealand, Panama, Paraguay, Peru, Singapore, South Korea, Taiwan, Thailand, Uruguay, and Venezuela.

Quantity

3,429 units, all models