Ossur Total Knee 2000, Model Number TK 2000, external prosthetic knee.

Recall: Z-1199-06 · Initiated March 29, 2006

Recall

Recall Number
Z-1199-06
Event Number
35132
Firm
Ossur North America Inc
FEI Number
1000519909
Product Code
ISY
Status
Terminated
Root Cause
Other
Initiated
March 29, 2006
Posted
July 4, 2006
Terminated
August 8, 2008
Address
27412 Aliso Viejo Pkwy, Aliso Viejo, CA, 92656-3371

Description

Ossur Total Knee 2000, Model Number TK 2000, external prosthetic knee.

Reason

The company's initiation of the recall is based on a finding that some units of the Total Knee device may contain faulty pins based in the axis of the knee. Small cracks propagate through the pin cross section until the pin breaks, severely compromising the knee function.

Action

On 3/29/06 a press release was posted on OandP.com, an industry specific web site. An announcement listing involved serial numbers and contact phone numbers was also posted on the company''s web site, ossur.com. A two stage mailing has been constructed. The first letter sent to all customers March 30, 2006, describes the problem and models affected. The second letter, March 31, 2006, is customer specific and lists all of the involved serial numbers by model which were purchased by that customer. Telephone calls will be placed to customers who do not respond to the mailings.

Distribution

Distribution worldwide in Austria, Canada, Columbia, Hong Kong, Japan, Korea, Kuwait, Malaysia, Mexico, Taiwan, & New Zealand. Involved states are: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, AND WY.

Quantity

1906