FDA Recall
Open, Classified
Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 127 cm (50"), REF US502, to prevent or treat hypothermia
Recall: Z-1198-2025
·
Initiated December 16, 2024
Recall
- Recall Number
- Z-1198-2025
- Event Number
- 96155
- Firm
- Augustine Temperature Management, LLC
- FEI Number
- 3005857264
- Product Code
- DWJ
- Status
- Open, Classified
- Root Cause
- Device Design
- Initiated
- December 16, 2024
- Posted
- February 20, 2025
- Address
- 15305 Minnetonka Blvd, Minnetonka, MN, 55345-1512
Description
Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 127 cm (50"), REF US502, to prevent or treat hypothermia
Reason
There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
Action
Augustine Temperature Management began notifying its consignees on 12/16/2024 via email and visit to exchange the affected product.
Distribution
US Nationwide distribution in the states of TX, AL, NV; AU, NZ.
Quantity
1 unit