FDA Recall Terminated

Heartstart HS1 Defibrillator Model M5068A

Recall: Z-1196-2010 · Initiated November 2, 2009

Recall

Recall Number
Z-1196-2010
Event Number
53816
Firm
Philips Medical Systems
FEI Number
1000524572
Product Code
NSA
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
November 2, 2009
Posted
March 23, 2010
Terminated
May 23, 2013
Address
2301 5th Ave, Ste 200, Seattle, WA, 98121

Description

Heartstart HS1 Defibrillator Model M5068A

Reason

Potential for failure of a capacitor during use which would prevent the AED from delivering effective defibrillation therapy when indicated.

Action

The firm notified their US distributors of the recall and instructed them to contact customers and provide information how to return the product to Philips for a free replacement. Firm contacted distributors via phone and email on 11/2/09 and by letter on 11/9/09. A web site providing information related to the recall was activated on 11/3/09.

Distribution

Worldwide Distribution.

Quantity

34 units distributed domestically