FDA Recall Terminated

Varian On-Board Imager Device versions 1.2 and 1.3, when used with Cone Beam Computed Tomography. Medical Linear Accelerator.

Recall: Z-1195-2007 · Initiated April 18, 2007

Recall

Recall Number
Z-1195-2007
Event Number
38349
Firm
Varian Medical Systems Inc
FEI Number
2916710
Product Code
IYE
Status
Terminated
Root Cause
Device Design
Initiated
April 18, 2007
Posted
August 29, 2007
Terminated
August 12, 2010
Address
911 Hansen Way, Palo Alto, CA, 94304

Description

Varian On-Board Imager Device versions 1.2 and 1.3, when used with Cone Beam Computed Tomography. Medical Linear Accelerator.

Reason

Patient positioning errors may occur because of mechanical variations during product installation.

Action

The firm issued a notification letter to all customers April 18, 2007. The firm intends to arrange for service personnel to contact affected sites to verify the calibration of the OBI system, and make arrangements to correct any issues. The firm will also update installation manuals to clearly describe the OBI installation and calibration requirements.

Distribution

Worldwide, including USA, Puerto Rico, Australia, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Finland, France, Germany, India, Ireland, Italy, Japan, Luxembourg, New Zealand, Norway, Poland, Portugal, Spain, Sweden, Switzerland, Netherlands, and United Kingdom.

Quantity

291 units