Medtronic Custom Pack EVNP Clinical with AP40 1/B (PERFUSION M448036B 1/B). The Affinity Pixie Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface is a single-use device designed to collect and store blood during extracorporeal circulation. Cardiotomy blood is collected, filtered, and defoamed before mixing with the venous blood, which is filtered. The primary blood-contacting surfaces of the CVR are coated with Balance Biosurface to reduce platelet activation and adhesion and preserve platelet function. (Generic Name: Cardiopulmonary bypass oxygenator).
Recall
- Recall Number
- Z-1194-2021
- Event Number
- 87271
- Firm
- Medtronic Perfusion Systems
- FEI Number
- 1000116158
- Product Code
- DTN
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- February 4, 2021
- Posted
- March 17, 2021
- Terminated
- April 26, 2024
- Address
- 7611 Northland Dr N, Brooklyn Park, MN, 55428-1088
Description
Medtronic Custom Pack EVNP Clinical with AP40 1/B (PERFUSION M448036B 1/B). The Affinity Pixie Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface is a single-use device designed to collect and store blood during extracorporeal circulation. Cardiotomy blood is collected, filtered, and defoamed before mixing with the venous blood, which is filtered. The primary blood-contacting surfaces of the CVR are coated with Balance Biosurface to reduce platelet activation and adhesion and preserve platelet function. (Generic Name: Cardiopulmonary bypass oxygenator).
Products being recalled due to potentially elevated level of bacterial endotoxin. Patients exposed to elevated levels of bacterial endotoxins may develop an acute inflammatory response.
A written Medical Device Correction letter will be mailed via UPS 2-day delivery to consignees who have received affected product according to Medtronic records. Consignees will be asked to complete the enclosed Customer Confirmation Form and email to [email protected]. Consignees will also be asked to return all unused affected product in your inventory to Medtronic. Contact Medtronic Customer Service at800-854-3570 to initiate a product return.
Worldwide distribution: US nationwide, Afghanistan, Argentina, Chile, China, Colombia, Egypt, Germany, Iraq, Italy, Japan, Korea, Kyrgyzstan, Latvia, Nicaragua, Paraguay, Poland, Qatar, Romania, Tanzania, United Arab Emirates, United Kingdom, Uzbekistan, and Viet Nam
10 units