FDA Recall Terminated

Stryker VertePort, 10 G Access Canula (short), REF 306-430, Qty 6; For use with the Stryker PCD Precision System and ACM Delivery System, Rx only, sterile, Stryker Instruments, Kalamazoo, MI.

Recall: Z-1194-2008 · Initiated February 13, 2008

Recall

Recall Number
Z-1194-2008
Event Number
46726
Firm
Stryker Instruments Div. of Stryker Corporation
FEI Number
1811755
Product Code
JDZ
Status
Terminated
Root Cause
Labeling Change Control
Initiated
February 13, 2008
Posted
March 11, 2008
Terminated
December 30, 2008
Address
4100 E Milham Ave, Portage, MI, 49002-9704

Description

Stryker VertePort, 10 G Access Canula (short), REF 306-430, Qty 6; For use with the Stryker PCD Precision System and ACM Delivery System, Rx only, sterile, Stryker Instruments, Kalamazoo, MI.

Reason

Mis-labeling; The instructions for use include a use for which the device is not intended. (Device not approved to capture a bone biopsy)

Action

Consignees were notified via Medical Device Recall Notification letter dated 2/13/08 to read the enclosed instructions for use, remove all old instructions for use at the consignee location and replace them with the new instructions for use, and to discard the old instructions for use.

Distribution

Nationwide.

Quantity

953 boxes Recalls Z-1192-1195-2008