FDA Recall
Terminated
Stryker VertePort, 10 G Access Canula (short), REF 306-430, Qty 6; For use with the Stryker PCD Precision System and ACM Delivery System, Rx only, sterile, Stryker Instruments, Kalamazoo, MI.
Recall: Z-1194-2008
·
Initiated February 13, 2008
Recall
- Recall Number
- Z-1194-2008
- Event Number
- 46726
- Firm
- Stryker Instruments Div. of Stryker Corporation
- FEI Number
- 1811755
- Product Code
- JDZ
- Status
- Terminated
- Root Cause
- Labeling Change Control
- Initiated
- February 13, 2008
- Posted
- March 11, 2008
- Terminated
- December 30, 2008
- Address
- 4100 E Milham Ave, Portage, MI, 49002-9704
Description
Stryker VertePort, 10 G Access Canula (short), REF 306-430, Qty 6; For use with the Stryker PCD Precision System and ACM Delivery System, Rx only, sterile, Stryker Instruments, Kalamazoo, MI.
Reason
Mis-labeling; The instructions for use include a use for which the device is not intended. (Device not approved to capture a bone biopsy)
Action
Consignees were notified via Medical Device Recall Notification letter dated 2/13/08 to read the enclosed instructions for use, remove all old instructions for use at the consignee location and replace them with the new instructions for use, and to discard the old instructions for use.
Distribution
Nationwide.
Quantity
953 boxes Recalls Z-1192-1195-2008