FDA Recall Terminated

AOS Trochanteric Nail, 10mm x 20cm x 135, Product Part Number: 1042-200.

Recall: Z-1191-2010 · Initiated January 14, 2010

Recall

Recall Number
Z-1191-2010
Event Number
54538
Firm
Advanced Orthopaedic Solutions Inc
FEI Number
3009833755
Product Code
HSB
Status
Terminated
Root Cause
Process control
Initiated
January 14, 2010
Posted
March 22, 2010
Terminated
April 26, 2012
Address
386 Beech Ave, Torrance, CA, 90501

Description

AOS Trochanteric Nail, 10mm x 20cm x 135, Product Part Number: 1042-200.

Reason

The recall was initiated after Advanced Orthopaedic Solutions (AOS) became aware of a manufacturing defect involving the short Trochanteric Nail, where one piece was found to be incorrectly bent. If this issue were to recur, surgery time could be extended, but only slightly as nail exchanges take place commonly in a surgery of this type.

Action

The recall was initiated with the firm forwarding a recall notice with attached customer response form on January 14, 2010 to the affected consignees. The recall notice informed the consignees of the reason for recall and gave them three steps to follow to inspect the affected Short Trochanteric Nail products. Consignees were provided with product return instruction for any incorrectly bent products found during their inspection. Customers with questions are instructed to contact the firm at (310) 533-9966.

Distribution

Nationwide and Japan

Quantity

27