FDA Recall Terminated

Tapered Screw-Vent MTX, Dental Implant, sterile. Model TSV4B10.

Recall: Z-1190-2018 · Initiated November 29, 2017

Recall

Recall Number
Z-1190-2018
Event Number
78667
Firm
Zimmer Dental Inc
FEI Number
2023141
Product Code
DZE
Status
Terminated
Root Cause
Packaging process control
Initiated
November 29, 2017
Posted
March 5, 2018
Terminated
February 8, 2019
Address
1900 Aston Ave, Carlsbad, CA, 92008-7308

Description

Tapered Screw-Vent MTX, Dental Implant, sterile. Model TSV4B10.

Reason

Two lots of the Tapered Screw-Vent Implants, with different outer thread diameters, were inadvertently commingled during the manufacturing process and may be mislabeled.

Action

On 11/29/2017, customers were notified via letter. Instructions for distributors include to ensure that affected team members are aware of the recall, immediately locate and quarantine affected devices in inventory, notify consignees if product was distributed, make arrangements to immediately return affected devices, and complete and return the Attachment 1 - Inventory Return Certification Form. Instructions for clinicians include to review the recall notice, examine inventory for affected devices, complete and return the Attachment 2 - Certificate of Acknowledgement, and return all affected devices.

Distribution

Worldwide distribution US nationwide, Canada, Chile, Costa Rica, Denmark, France, Hungary, Italy, Japan, Lebanon, Portugal, Serbia, Spain, and Taiwan.

Quantity

995 distribution events total