FDA Recall Terminated

Volcano s5 Imaging system, Ultrasound pulsed echo imaging system. Product consists of the CPU, control console, monitor, printer, and patient interface module (PIM)

Recall: Z-1190-2007 · Initiated April 2, 2007

Recall

Recall Number
Z-1190-2007
Event Number
38324
Firm
Volcano Corporation
FEI Number
2939520
Product Code
IYO
Status
Terminated
Root Cause
Other
Initiated
April 2, 2007
Posted
August 23, 2007
Terminated
June 3, 2008
Address
2870 Kilgore Rd, Rancho Cordova, CA, 95670-6133

Description

Volcano s5 Imaging system, Ultrasound pulsed echo imaging system. Product consists of the CPU, control console, monitor, printer, and patient interface module (PIM)

Reason

During emissions testing, it was discovered that the product exceeds emissions limits for electromagnetic compatibility.

Action

The firm notified its customers by letter beginning April 2, 2007, and started replacing the monitor with appropriate replacement parts.

Distribution

Product was distributed to health care facilities nationwide and to distributors in Europe (via subsidiary), Japan (via subsidiary), Latin America and Asia (third party distributors).

Quantity

351 units worldwide