FDA Recall
Terminated
Volcano s5 Imaging system, Ultrasound pulsed echo imaging system. Product consists of the CPU, control console, monitor, printer, and patient interface module (PIM)
Recall: Z-1190-2007
·
Initiated April 2, 2007
Recall
- Recall Number
- Z-1190-2007
- Event Number
- 38324
- Firm
- Volcano Corporation
- FEI Number
- 2939520
- Product Code
- IYO
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 2, 2007
- Posted
- August 23, 2007
- Terminated
- June 3, 2008
- Address
- 2870 Kilgore Rd, Rancho Cordova, CA, 95670-6133
Description
Volcano s5 Imaging system, Ultrasound pulsed echo imaging system. Product consists of the CPU, control console, monitor, printer, and patient interface module (PIM)
Reason
During emissions testing, it was discovered that the product exceeds emissions limits for electromagnetic compatibility.
Action
The firm notified its customers by letter beginning April 2, 2007, and started replacing the monitor with appropriate replacement parts.
Distribution
Product was distributed to health care facilities nationwide and to distributors in Europe (via subsidiary), Japan (via subsidiary), Latin America and Asia (third party distributors).
Quantity
351 units worldwide