Vapofils with Stainless Steel Block Ends, Item Number V0507S (Sevofluane). The vapofil is a keyed anesthetic transfer tube that transfers the anesthetic from the bottle to the vaporizer.
Recall
- Recall Number
- Z-1190-06
- Event Number
- 35358
- Firm
- Southmedic Inc Po Box 1365 Barrie Canada Ontario
- FEI Number
- 3001717360
- Product Code
- CAD
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 21, 2006
- Posted
- July 4, 2006
- Terminated
- March 6, 2012
Description
Vapofils with Stainless Steel Block Ends, Item Number V0507S (Sevofluane). The vapofil is a keyed anesthetic transfer tube that transfers the anesthetic from the bottle to the vaporizer.
Southmedic has initiated this recall because their Japanese distributor received field reports of leaking vapofils. Leakage may cause non-toxic spills in the hospital environment, distraction of practitioners during conduct of anesthesia and surgery, and may damage certain plastic equipment. Leakage may also expose patients to low concentrations of volatile anesthetics.
Recall letters were sent to both US consignees via US Postal Service delivery with signature receipt required on April 21, 2006. Each letter included a Recall and Effectiveness Check Form for each consignee to fill out and fax to Southmedic. U.S. and international consignees were instructed to examine inventory immediately to identify if they have any product on hand, discontinue the distribution of these products and return all affected products to Southmedic. Consignees were also instructed to immediately notify their customers if they have further distributed these products.
US in FL & NJ Worldwide in Australia, Belgium, Ecuador, Germany, Japan, Netherlands, Sri Lanka, Taiwan, & United Arab Emerates
186 in US (403 Worldwide)