FDA Recall Open, Classified

Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 10.1.0.613 and 10.1.1.54 Software Version: RayStation 10B and 10B Service Pack 1

Recall: Z-1188-2025 · Initiated January 16, 2025

Recall

Recall Number
Z-1188-2025
Event Number
96156
Firm
RAYSEARCH LABORATORIES AB Eugeniavagen 18c Stockholm Sweden
FEI Number
3007774465
Product Code
MUJ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
January 16, 2025
Posted
February 19, 2025

Description

Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 10.1.0.613 and 10.1.1.54 Software Version: RayStation 10B and 10B Service Pack 1

Reason

Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

Action

Raysearch issued Field Safety Notice, Medical Device Correction #148655 RayStation version 4.5 to 2024B including some service packs via email on 1/17/25. Letter states reason for recall, health risk and action to take: For users of HU-to-mass density CT calibration curves If a user wants to evaluate a plan with respect to relative stopping power/WE range error rather than mass density errors, there are two options: Option 1. Use a HU-to-SPR CT calibration curve instead of a HU-to-mass density curve. " For evaluation of an existing case, the evaluation can be done on an anonymized and exported copy of the patient and plan. " Contact RaySearch Service if you need help with the creation of a HU-to-SPR curve from an existing HU-to-mass density curve. Option 2. Evaluate the plan using a higher (effective) mass density uncertainty that gives similar relative stopping power and WE range change as the nominal value. " The following values of Effective mass density errors that give similar result as an intended Stopping power error have been deduced from a limited set of patients: " For other values of intended Stopping power error, interpolate the approximate Effective mass density value from the table above. " The values are estimates based on a few patients and are expected to vary slightly depending on the mass densities along the beam path. " Note that the values are not symmetrical around 0. For users of HU-to-SPR CT calibration curves Be aware that the density uncertainty given in Robust evaluation, Compute perturbed dose and Robust optimization is used directly to scale stopping power and therefore WE range. Starting from RayStation v2025, scheduled for market release in April 2025 (subject to market clearance in some markets), the impact of the mass density uncertainty in Robust evaluation and Compute perturbed dose when using a HU-to-mass density CT calibration curve will be changed to harmonize with the other use cases. This means that as of RayStat

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Belgium ENG, Canada, China, China APactron, Denmark, France, Germany, Hong Kong S.A.R. China, India, Italy, Japan, Korea, Netherlands, Norway, Poland, Singapore, Spain, SwitzerlandENG, Taiwan, Thailand, UK.

Quantity

0