FDA Recall Terminated

HAMILTON G5 ventilators with software version V2.0x. Hamilton Medical AG, Switzerland Imported and distributed by Hamilton Medical, 4990 Energy Way, Reno, NV 89502. Product Usage: Hamilton-G5 Ventilators are designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The devices are intended for use in the hospital and institutional environment with trained healthcare professionals.

Recall: Z-1185-2012 · Initiated February 6, 2012

Recall

Recall Number
Z-1185-2012
Event Number
61238
Firm
Hamilton Medical Inc
FEI Number
2937708
Product Code
CBK
Status
Terminated
Root Cause
Software design
Initiated
February 6, 2012
Posted
March 7, 2012
Terminated
October 23, 2013
Address
4990 Energy Way P O Box 30008, Reno, NV, 89502-3008

Description

HAMILTON G5 ventilators with software version V2.0x. Hamilton Medical AG, Switzerland Imported and distributed by Hamilton Medical, 4990 Energy Way, Reno, NV 89502. Product Usage: Hamilton-G5 Ventilators are designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The devices are intended for use in the hospital and institutional environment with trained healthcare professionals.

Reason

Hamilton-G5 ventilators with software version V2.0x may experience an unintentional change of the displayed ventilation mode under certain conditions.

Action

Hamilton Medical sent a Medical Device Safety Alert and Correction Action letter dated February 17, 2011 to all affected customers via a traceable method. The letter identified the affected product, problem and immediate actions required by operators, distributor and manufacturer. The letter states that a Hamilton Medical Account manager will contact customers to schedule a convenient time for the Technical Support Department to install the software update. Questions or concerns may be directed to Technical Support Manager at 800-426-6331, Ext 125.

Distribution

(USA) Nationwide Distribution including the states of: CA, CO, FL, GA, MA, MI, MO, NC, NV, NY, PA, TN, TX, VA and WV.

Quantity

456 devices in US.