FDA Recall Terminated

Access Power Supply Assembly Sled Part Number: A48196/6845C; Component of: Access and Access 2 Immunoassay Systems, SYNCHRON LXi 725 System, and UniCel DxC 600i System, for in vitro diagnostic use.

Recall: Z-1183-2008 · Initiated October 17, 2007

Recall

Recall Number
Z-1183-2008
Event Number
46316
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
JJE
Status
Terminated
Root Cause
Process control
Initiated
October 17, 2007
Posted
August 20, 2008
Terminated
April 27, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

Access Power Supply Assembly Sled Part Number: A48196/6845C; Component of: Access and Access 2 Immunoassay Systems, SYNCHRON LXi 725 System, and UniCel DxC 600i System, for in vitro diagnostic use.

Reason

Electrical grounding failure: During manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering.

Action

Beckman Coulter mailed a Product Corrective Action (PCA) letter on October 17, 2007 to all Access and Access 2 lmmunoassay Systems, SYNCHRON LXi 725 System, and UniCel DxC 600i System customers. The customers were informed that during manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering. The letter referenced a proactive initiative to inspect the power supply sled assemblies on all instruments referenced. A response form was included.

Distribution

Nationwide Distribution and Canada

Quantity

3055 instruments in the US; 192 instruments in Canada