FDA Recall Terminated

McKesson Horizon Medical Imaging (HMI) products versions 4.6.1 to including 11.9 and McKesson Radiology (MR) products 12.0 and 12.1.1.

Recall: Z-1182-2016 · Initiated February 18, 2016

Recall

Recall Number
Z-1182-2016
Event Number
73452
Firm
Mckesson Medical Immaging Suite 130-10711 Cambie Road Richmond Canada British Columbia
FEI Number
3003757840
Product Code
LLZ
Status
Terminated
Root Cause
Software design
Initiated
February 18, 2016
Terminated
August 28, 2018

Description

McKesson Horizon Medical Imaging (HMI) products versions 4.6.1 to including 11.9 and McKesson Radiology (MR) products 12.0 and 12.1.1.

Reason

McKesson has identified a design deficiency where under rare circumstances, imported images/studies may re-use a non-unique image directory. This issue may cause incorrect images to be displayed for a patient.

Action

McKesson sent an Urgent Field Safety Notice dated February 18, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Immediate actions to take to reduce the risk to patient All customers receiving this safety notice must contact McKesson Support as soon as possible at 1-800-663-2533 US/Canada and International Toll Free Radiology - 00 800 626 20009: 1) To acknowledge that they have read and understood this Field Safety Notice And 2) Arrange installation of software update/upgrade. This notice should be distributed to all personnel within your organization who need to be aware of this issue and the required actions, and alert other affiliated parties that are affected by this Field Safety Notice. For further questions, please call (404) 338-3556

Distribution

Worldwide Distribution - USA (nationwide) including Puerto Rico and to the states of : AL, AK, AZ, AR, CA, CO, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OR, PA, RI, SC, TN, TX, VT, VA, WA, WV, WI, WY., and the countries of : Guam, Great Britain, France, Canada, Ireland and Australia

Quantity

631 devices