FDA Recall Terminated

Intellivue Central Station Software for the: Intellivue Information Center; Philips Information Center, Central Station Models: M3145, M3150, M3154, M3155, M3169

Recall: Z-1182-2007 · Initiated July 16, 2007

Recall

Recall Number
Z-1182-2007
Event Number
38400
Firm
Philips Medical Systems
FEI Number
1218950
Product Code
MHX
Status
Terminated
Root Cause
Other
Initiated
July 16, 2007
Posted
September 5, 2007
Terminated
August 1, 2016
Address
3000 Minuteman Road, Andover, MA, 01810

Description

Intellivue Central Station Software for the: Intellivue Information Center; Philips Information Center, Central Station Models: M3145, M3150, M3154, M3155, M3169

Reason

Incorrect Patient Assignment: the information center will in most cases be associated with the "incorrect" patient. This occurs on both standalone and networked systems running fixed mode monitoring. (This does not occur on systems that run flex mode monitoring)

Action

Philips Medical Systems issued a letter notification dated July 2007 " Urgent Device Correction". Customers with networked information centers will be corrected through a database server. Stand-Alone units will have a software correction. Software version J.00.25 will correct the problem. Users are advised not to use th 12 Lead Analysis and Export Option in fixed or flex mode monitoring unitl the software correction is made.

Distribution

Worldwide Distribution: USA including states of NJ, NY, OK, MI, and UT and Foreign: Australia, Finland, France, Netherlands, and the UK

Quantity

76 units