Intellivue Central Station Software for the: Intellivue Information Center; Philips Information Center, Central Station Models: M3145, M3150, M3154, M3155, M3169
Recall
- Recall Number
- Z-1182-2007
- Event Number
- 38400
- Firm
- Philips Medical Systems
- FEI Number
- 1218950
- Product Code
- MHX
- Status
- Terminated
- Root Cause
- Other
- Initiated
- July 16, 2007
- Posted
- September 5, 2007
- Terminated
- August 1, 2016
- Address
- 3000 Minuteman Road, Andover, MA, 01810
Description
Intellivue Central Station Software for the: Intellivue Information Center; Philips Information Center, Central Station Models: M3145, M3150, M3154, M3155, M3169
Incorrect Patient Assignment: the information center will in most cases be associated with the "incorrect" patient. This occurs on both standalone and networked systems running fixed mode monitoring. (This does not occur on systems that run flex mode monitoring)
Philips Medical Systems issued a letter notification dated July 2007 " Urgent Device Correction". Customers with networked information centers will be corrected through a database server. Stand-Alone units will have a software correction. Software version J.00.25 will correct the problem. Users are advised not to use th 12 Lead Analysis and Export Option in fixed or flex mode monitoring unitl the software correction is made.
Worldwide Distribution: USA including states of NJ, NY, OK, MI, and UT and Foreign: Australia, Finland, France, Netherlands, and the UK
76 units