FDA Recall Terminated

Quick Connects. Steris Corporation. Model numbers: QKC1736E, QKC1736INT, QKC1737E, and QKC1737INT. Each Quick Connect contains a flow unit consisting of tubing, tethers, and adapters. For sterilant and rinse water used in endoscope washers.

Recall: Z-1180-2016 · Initiated December 4, 2015

Recall

Recall Number
Z-1180-2016
Event Number
73009
Firm
Steris Corporation
FEI Number
1527821
Product Code
MED
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
December 4, 2015
Terminated
August 26, 2016
Address
5960 Heisley Rd, Mentor, OH, 44060-1834

Description

Quick Connects. Steris Corporation. Model numbers: QKC1736E, QKC1736INT, QKC1737E, and QKC1737INT. Each Quick Connect contains a flow unit consisting of tubing, tethers, and adapters. For sterilant and rinse water used in endoscope washers.

Reason

The firm discovered that one of the hysteroscope models listed for processing in four existing Quick Connects has one inlet port rather than two inlet ports as indicated in the Quick Connect processing instructions.

Action

On 1/18/2016 the firm sent Voluntary Field Correction NOtices to their customers.

Distribution

Worldwide distribution. US nationwide, Australia, Bahrain, Bulgaria, Canada, Greece, Italy, Malaysia, Mexico, Qatar, Russia, Spain, and United Kingdom.

Quantity

809 units