FDA Recall Terminated

Pelton & Crane Dental Chairs, Spirit 1500 with magnetic headrest, Pelton & Crane, Charlotte, NC 28273. Intended for the properly positioning of patients to perform dental procedures.

Recall: Z-1180-2010 · Initiated February 24, 2010

Recall

Recall Number
Z-1180-2010
Event Number
54719
Firm
Pelton & Crane Company
FEI Number
1017522
Product Code
KLC
Status
Terminated
Root Cause
Device Design
Initiated
February 24, 2010
Posted
April 1, 2010
Terminated
October 28, 2010
Address
11727 Fruehauf Dr, Charlotte, NC, 28273

Description

Pelton & Crane Dental Chairs, Spirit 1500 with magnetic headrest, Pelton & Crane, Charlotte, NC 28273. Intended for the properly positioning of patients to perform dental procedures.

Reason

Magnetic headrests may affect patients with implantable pacemakers and defibrillators.

Action

Pelton & Crane Issued an "Urgent Field Safety Alert" notification dated February 24, 2010. Consignees were instructed of the affected product and informed of proper actions to take. For further information, contact Pelton & Crane at 1-704-588-2126.

Distribution

Worldwide Distribution -- United States, Australia, the Netherlands, Canada and Germany.

Quantity

Total: 498 chairs and 619 replacement upholstery spare parts