FDA Recall Terminated

Theratron 780E Cobalt Radiotherapy

Recall: Z-1179-06 · Initiated July 8, 2004

Recall

Recall Number
Z-1179-06
Event Number
35617
Firm
MDS Canada Inc. DBA MDS Nordion
FEI Number
3001092088
Product Code
IWB
Status
Terminated
Root Cause
Other
Initiated
July 8, 2004
Posted
June 27, 2006
Terminated
June 27, 2006
Address
447 March Road, Kanata Canada

Description

Theratron 780E Cobalt Radiotherapy

Reason

The source locking device may not fit properly through the smaller opening in the new field light shield which could prevent the source from being returned manually under emergency conditions.

Action

A User Bulletin (CUB-2004-01dated 7/5/04) was issued to affected users on 8/24/04 and 8/31/04. The bulletin provided 2 options for corrective action.

Distribution

The products were shipped to medical facilities in FL and KS. The product was also shipped to Kuwait, Spain, Honduras, Romania, France, Grrece, Bangladesh, Germany, Nepal, Canada, England, Thailand, Phillippines, Mexico, Columbia, India, Switzerland, Syria, Trinidad & Tobago, Ireland, Turkey, Algeria, Portugal, Poland, Iran, China, Cuba, Belgium, Hungary, Egypt, Italy, Vietnam, Sri Lanka, and Costa Rica.

Quantity

78 units