FDA Recall
Terminated
Roche Calcium Reagent Kit for use on the Cobas MIRA Chemistry System; in vitro diagnostic; Cat. No. 04585984160.
Recall: Z-1176-06
·
Initiated June 7, 2006
Recall
- Recall Number
- Z-1176-06
- Event Number
- 35493
- Firm
- Roche Diagnostics Corp.
- FEI Number
- 1823260
- Product Code
- CIC
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 7, 2006
- Posted
- June 27, 2006
- Terminated
- April 23, 2008
- Address
- 9115 Hague Rd, Indianapolis, IN, 46256-1025
Description
Roche Calcium Reagent Kit for use on the Cobas MIRA Chemistry System; in vitro diagnostic; Cat. No. 04585984160.
Reason
Falsely low patient and control results may be reported when the reagent has been on-board for less than the 8 hour use period specified in the labeling.
Action
Consignees were notified via letter dated 6/6/06 and sent via UPS to discontinue use of these lots and to destroy them. Distributors are being asked to notify their customers.
Distribution
Nationwide.
Quantity
933 kits