Novafil Monofilament Polybutester Suture Product Description: 8886 4002-41 NOVAFIL 3-0 BLU 90CM C14 8886 4400-13 NOVAFIL 6-0 BLU 45CM SBE2 8886 4400-23 NOVAFIL 5-0 BLU 45CM SBE2 8886 440123 NOVAFIL 5-0 BLU 45CM SBE3 8886 4402-33 NOVAFIL 4-0 BLU 45CM SBE4 8886 4403-33 NOVAFIL 4-0 BLU 45CM SBE6 8886 4410-13 NOVAFIL 6-0 BLU 45CM P10 8886 4410-03 NOVAFIL 7-0 BLU 45CM P10 8886 4422-43 NOVAFIL 3-0 BLU 45CM C14 8886 4424-51 NOVAFIL 2-0 BLU 75CM C16 8886 4450-41 NOVAFIL 3-0 BLU 75CM V26 8886 4452-51 NOVAFIL 2-0 BLU 75CM V20 8886 4459-61 NOVAFIL 0 BLU 75CM GS22 8886 4470-81 NOVAFIL* 2 BLU 150CM GS26 8886 4582-53 NOVAFIL* 2-0 BLU 75CM P14 SPB-1213G NOVAFIL* 6-0 BLU 45CM P13X12 SPB-1233G NOVAFIL* 4-0 BLU 45CM P13X12 SPB-1623G NOVAFIL* 5-0 BLU 45CM P12X12 SPB-5142G NOVAFIL* 5-0 BLU 45CM P11X12 SPB-5143G NOVAFIL* 4-0 BLU 45CM P11X12 SUT SPB-5223G NOVAFIL 5-0 BLU 45CM P13 SPB-5433G NOVAFIL* 4-0 CLR 45CM P24X12 SPB-5633G NOVAFIL* 4-0 BLU 45CM P12X12
Recall
- Recall Number
- Z-1175-2023
- Event Number
- 91324
- FEI Number
- 1219930
- Product Code
- GAT
- Status
- Open, Classified
- Root Cause
- Packaging process control
- Initiated
- December 2, 2022
- Posted
- February 25, 2023
- Address
- LP 60 Middletown Ave, Covidien, North Haven, CT, 06473-3908
Description
Novafil Monofilament Polybutester Suture Product Description: 8886 4002-41 NOVAFIL 3-0 BLU 90CM C14 8886 4400-13 NOVAFIL 6-0 BLU 45CM SBE2 8886 4400-23 NOVAFIL 5-0 BLU 45CM SBE2 8886 440123 NOVAFIL 5-0 BLU 45CM SBE3 8886 4402-33 NOVAFIL 4-0 BLU 45CM SBE4 8886 4403-33 NOVAFIL 4-0 BLU 45CM SBE6 8886 4410-13 NOVAFIL 6-0 BLU 45CM P10 8886 4410-03 NOVAFIL 7-0 BLU 45CM P10 8886 4422-43 NOVAFIL 3-0 BLU 45CM C14 8886 4424-51 NOVAFIL 2-0 BLU 75CM C16 8886 4450-41 NOVAFIL 3-0 BLU 75CM V26 8886 4452-51 NOVAFIL 2-0 BLU 75CM V20 8886 4459-61 NOVAFIL 0 BLU 75CM GS22 8886 4470-81 NOVAFIL* 2 BLU 150CM GS26 8886 4582-53 NOVAFIL* 2-0 BLU 75CM P14 SPB-1213G NOVAFIL* 6-0 BLU 45CM P13X12 SPB-1233G NOVAFIL* 4-0 BLU 45CM P13X12 SPB-1623G NOVAFIL* 5-0 BLU 45CM P12X12 SPB-5142G NOVAFIL* 5-0 BLU 45CM P11X12 SPB-5143G NOVAFIL* 4-0 BLU 45CM P11X12 SUT SPB-5223G NOVAFIL 5-0 BLU 45CM P13 SPB-5433G NOVAFIL* 4-0 CLR 45CM P24X12 SPB-5633G NOVAFIL* 4-0 BLU 45CM P12X12
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
Medtronic issued an "URGENT: MEDICAL DEVICE RECALL" letter to US consignees on 01-Dec-2022 via UPS 2-day delivery. OUS countries, notifications are handled via a regionally approved method. Letter states reason for recall, health risk and actions to take: Actions: " Identify and quarantine all unused and non-expired affected non-absorbable sutures. " Please note the affected product may be located within Custom Suture Packs. Please reference Custom Suture Packs to identify affected product. " Please complete the enclosed Customer Confirmation Form (even if your account no longer has affected product on hand) and email to [email protected].> " Return all unused and non-expired affected product in your inventory to Medtronic as indicated in the Shipping and Return Instructions below. " Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected product has been transferred or distributed. Shipping and Return Instructions: " Product purchased directly from Medtronic please contact [email protected] for Return Good Authorization (RGA). Your local Medtronic Representative can assist you as necessary in initiating the return of this product. " Credit or Replacement for the returned affected product will be issued based on the RGA number. " Return all unused and non-expired affected product(s) in your inventory to: Medtronic, Attn: Field Returns Dept. 195 McDermott Road, North Haven, CT 06473 USA. " If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. If you have any questions regarding this communication, please contact your Medtronic Representative or Customer Service at 800-962-9888.
US Nationwide Distribution and OUS Foreign distribution to countries of: Albania Argentina Armenia Australia Austria Azerbaijan Belarus Belgium Bosnia And Herzegovina Brazil Bulgaria Cambodia Cameroon Canada Canary Islands Chile China Colombia Croatia Czech Republic Denmark Egypt Finland France Georgia Germany Ghana Greece Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy JAPAN Jordan Kazakhstan Kenya Korea, Republic Of Kosovo Kuwait Latvia Lebanon Lithuania Macao Malaysia Malta Mauritius Mexico Moldova, Republic Of Morocco Myanmar Namibia Netherlands New Zealand North Macedonia Norway Pakistan Panama Paraguay Philippines Poland Portugal Reunion Romania Russian Federation Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Syrian Arab Republic Taiwan Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Uruguay Uzbekistan Viet Nam Zimbabwe
48492 units