FDA Recall Terminated

ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-7 Taper Needles, Catalog Number: 223114. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.

Recall: Z-1174-2013 · Initiated March 22, 2013

Recall

Recall Number
Z-1174-2013
Event Number
64706
Firm
DePuy Mitek, Inc., a Johnson & Johnson Co.
FEI Number
1221934
Product Code
NEW
Status
Terminated
Root Cause
Labeling Change Control
Initiated
March 22, 2013
Posted
April 25, 2013
Terminated
February 24, 2014
Address
325 Paramount Drive, Raynham, MA, 02767-5199

Description

ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-7 Taper Needles, Catalog Number: 223114. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.

Reason

The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six

Action

On March 22, 2013, firm notified consignees of recall via letter. Consignees were informed of the following: The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six weeks after implantation. As Blue and Violet ORTHOCORD free strand sutures share the same IFU, IFU-108147 will be revised. Consignees were requested to confirm the receipt of the letter by March 29, 2013.

Distribution

Nationwide Distribution.

Quantity

206540