FDA Recall Terminated

Cook brand Zenith AAA Bifurcated main body graft with the H&L-B One-Shot Introducer System, preloaded bifurcated main body graft - aortic graft diameter 24 mm - iliac leg diameters 11 mm - main body lengths cntralateral limb 103mm ipsilateral limb 133 mm - delivery system 18 FR; Catalog no. TFB-24-103.

Recall: Z-1166-04 · Initiated June 10, 2004

Recall

Recall Number
Z-1166-04
Event Number
29300
Firm
Cook, Inc.
FEI Number
1820334
Product Code
MIH
Status
Terminated
Root Cause
Other
Initiated
June 10, 2004
Posted
July 27, 2004
Terminated
November 1, 2004
Address
750 Daniels Way, Bloomington, IN, 47404-9120

Description

Cook brand Zenith AAA Bifurcated main body graft with the H&L-B One-Shot Introducer System, preloaded bifurcated main body graft - aortic graft diameter 24 mm - iliac leg diameters 11 mm - main body lengths cntralateral limb 103mm ipsilateral limb 133 mm - delivery system 18 FR; Catalog no. TFB-24-103.

Reason

Introducer tubing may contain a vein of unextruded material and carbon black or a strand of tubing material may be displaced from the introducer during use.

Action

The firm issued recall letters to customers on 6/10/04. The recall was extended and URGENT EXTENSION OF PRODUCT RECALL notices were sent out on June 23, 2004.

Distribution

Nationwide, Canada, Denmark and Japan.

Quantity

45