FDA Recall Terminated

Oxford Partial Knee System Right Medial Tibial Trial Tray Size C Product Usage: Instruments for use with the Oxford Uni Partial Knee System

Recall: Z-1164-2017 · Initiated January 5, 2016

Recall

Recall Number
Z-1164-2017
Event Number
76179
Firm
Biomet U.K., Ltd. Waterton Industrial Estate Bridgend South Wales United Kingdom
FEI Number
3002806535
Product Code
NRA
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
January 5, 2016
Terminated
June 30, 2018

Description

Oxford Partial Knee System Right Medial Tibial Trial Tray Size C Product Usage: Instruments for use with the Oxford Uni Partial Knee System

Reason

Zimmer Biomet is conducting a medical device field action for various lot-specific Oxford Knee instruments due to potential alumina inclusions in the raw material batch used to produce the affected products. Inclusions contained within the finished product could lead to the cracking and separation of the instrument.

Action

Zimmer Biomet initiated a voluntary recall of the Oxford Knee instruments due to potential alumina inclusions in the raw material batch used to produce the ffected products by mailing letters via FedEx on 01/05/2017. Customers were instructed to Review this notification and ensure affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in their inventory. 3. Complete the Certification of Acknowledgement portion of Attachment 1 Inventory Return Certification Form. a. Return a digital copy to [email protected] within three (3) days. Immediately return all affected product from your distributorship and affected hospitals within your territory along with a completed Attachment 1 Inventory Return Certification Form and Attachment 2 Certificate of Decontamination to Zimmer Biomet. a. Request a Return Authorization Number via email to [email protected] or through FAST/SMS. Be sure to specify RECALL as the RGA type. b. For each return, send a copy of Attachment 1 and Attachment 3 to [email protected]. c. Include a hardcopy of Attachment 1 with your shipment for immediate processing. d. Include a hardcopy of Attachment 2 with returned instruments. Mark the outside of the returned boxes clearly with RECALL. 5. Using the Additional Accounts Form provided with the email notice sent to your facility, return contact information for any additional hospitals that may have received or used the affected product. If there are no additional users to notify, please indicate none or NA on the form and return. 6. Retain a copy of your recall acknowledgement and product return forms for your records in the event of a compliance audit of your facility. 7. If after reviewing the recall notice you have further questions or concerns please call the customer call center at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outs

Distribution

Worldwide Distribution in the states of LA, NC, AR, MT, MO, CA, NY, KS, TX, MA, KY, OK, SC, NJ and the countries of Foreign: THAILAND, JAPAN, NETHERLANDS

Quantity

72 units in total