FDA Recall
Terminated
Sunrise Medical Quickie Power Wheelchairs: Three models -Rhythm, Groove, and Zippy Z-Bop (with Quickie IQ Enhanced Displays)
Recall: Z-1163-2007
·
Initiated April 23, 2007
Recall
- Recall Number
- Z-1163-2007
- Event Number
- 38309
- Firm
- Sunrise Medical Inc
- FEI Number
- 2937137
- Product Code
- ITI
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- April 23, 2007
- Posted
- August 14, 2007
- Terminated
- March 25, 2008
- Address
- 2842 N Business Park Ave, Fresno, CA, 93727-1328
Description
Sunrise Medical Quickie Power Wheelchairs: Three models -Rhythm, Groove, and Zippy Z-Bop (with Quickie IQ Enhanced Displays)
Reason
The wire harness connected to the display housing (Quickie IQ Enhanced Displays) may separate and cause sparks from a short circuit.
Action
Firm contacted consignees by phone and mailed May 1, 2007, Urgent Medical Device notification letters.
Distribution
Product was distributed Worldwide.
Quantity
229 Enhanced Display units (affecting 220 wheelchairs)