FDA Recall Open, Classified

MOSAIQ Oncology Information System

Recall: Z-1162-2025 · Initiated January 15, 2025

Recall

Recall Number
Z-1162-2025
Event Number
96153
Firm
Elekta, Inc.
FEI Number
1037831
Product Code
IYE
Status
Open, Classified
Root Cause
Software design
Initiated
January 15, 2025
Posted
February 12, 2025
Address
400 Perimeter Center Ter NE, Ste 50, Atlanta, GA, 30346-1227

Description

MOSAIQ Oncology Information System

Reason

Image offsets calculated by oncology Information System software may be incorrect for Linear Accelerator (LINACs) not characterized at IEC61217 scale.

Action

On January 15, 2025 Elekta issued a recall notification to affected consignees via Email. Elekta asked consignees to take the following actions. 1. Customers included in the scope of the Field Safety Notice, should not calculate shifts from 2D verification images taken at a non-zero patient support angle (couch angle). 2. Please post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed. 3. Advise the appropriate personnel, working with this product, on the content of this letter. 4. You are required to either acknowledge receipt of this notification via the Elekta Care" Community or complete this form and return it to Elekta immediately upon receipt, but no later than within 30 days.

Distribution

US: PA, AL, FL, IA UT OUS: Austria Belgium Bolivia Bosnia and Herzegowina Brazil Canada France Germany Hong Kong Italy Japan Korea, Republic of Malta Netherlands Panama South Africa Spain Taiwan Thailand Turkey Venezuela

Quantity

64 units