FDA Recall Terminated

OBSERVA R02 (version R02.00.17) computers using PSC 6000 or HHP Barcode Scanners which are connected to BacT/ALERT 3D

Recall: Z-1161-06 · Initiated March 28, 2006

Recall

Recall Number
Z-1161-06
Event Number
35518
Firm
bioMerieux Inc
FEI Number
1950204
Product Code
MDB
Status
Terminated
Root Cause
Other
Initiated
March 28, 2006
Posted
June 22, 2006
Terminated
March 7, 2008
Address
595 Anglum Rd, Hazelwood, MO, 63042-2320

Description

OBSERVA R02 (version R02.00.17) computers using PSC 6000 or HHP Barcode Scanners which are connected to BacT/ALERT 3D

Reason

Software problem with scanner inter-character delay which may result in false positive or false negative results after incorrect scanning of bottle ID's

Action

The recalling firm telephoned customers who ordered the MB bottles and use the OBSERVA software ver. 2.0 to inform them of the issues. Recall letters were issued to all customers via overnight mail on 5/22/06 informing them a field representative would be visiting to make a short-term correction to the system. The letter also informed the customer a permanent correction to the software was scheduled for release later in the year.

Distribution

WorldwideDistribution--Nationwide to medical centers. Government distribution was made to MN and CA, and military distribution was made to OK. Foreign distribution was made to Thailand, The Netherlands, Germany, Greece, Portugal, and France.

Quantity

51 units