FDA Recall Terminated

Digitial Medica X-ray Imagining System, Model: uDR 596i, RX, CE0123, UDI:(01) 06971576833023 - Product Usage: intended to used by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, limbs and trunk. Not for mammography.

Recall: Z-1160-2021 · Initiated January 25, 2021

Recall

Recall Number
Z-1160-2021
Event Number
87289
Firm
Shanghai United Imaging Healthcare Co., Ltd. No. 2258 Chengbei Rd, Jiading Ind. Shanghai China
FEI Number
3011015597
Product Code
KPR
Status
Terminated
Root Cause
Device Design
Initiated
January 25, 2021
Terminated
May 3, 2022

Description

Digitial Medica X-ray Imagining System, Model: uDR 596i, RX, CE0123, UDI:(01) 06971576833023 - Product Usage: intended to used by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, limbs and trunk. Not for mammography.

Reason

There is a potential for the X-Ray tube head (Touch screen interface) to perform an uncommanded motion. This could result in a collision with the patient.

Action

On 01/25/2021, Shanghai United Imaging sent an Urgent: Medical Device Correction via: Email. In addition to informing consignees about the recall, the firm asked customers to take the following actions: 1. Once the RHA motor driver is powered on, and while the RHA motor is looking for the zero-position (before its first electric movements), please do not apply any external force to the TUI. 2. For further questions, please contact your United Imaging Service Engineer or call the United Imaging Customer Contact Center at our 24-hour hotline (855) 221-1552 (toll free) or (832) 699-2799 (direct). 3. Adverse reactions or quality problems experienced with the use of this product may be reported to FDAs MedWatch Adverse Event Reporting program either online, by regulatory mail or by fax. 4. Notify all users of the affected products via this Customer Notification Letter. 5. Please completed the Receipt of Customer Notification Letter form and return to United Imaging.

Distribution

US Nationwide distribution in the state of TX.

Quantity

6 units