FDA Recall Terminated

Smith & Nephew RENASYS-Gauze with Soft Port Kit REF #'s: 66800933, 66800934, 66800935, and 66800936 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)

Recall: Z-1160-2014 · Initiated January 9, 2014

Recall

Recall Number
Z-1160-2014
Event Number
67477
Firm
Smith & Nephew Inc.
FEI Number
3006760724
Product Code
OMP
Status
Terminated
Root Cause
Packaging change control
Initiated
January 9, 2014
Posted
March 10, 2014
Terminated
September 26, 2014
Address
970 Lake Carillon Dr, Ste 110, Saint Petersburg, FL, 33716-1130

Description

Smith & Nephew RENASYS-Gauze with Soft Port Kit REF #'s: 66800933, 66800934, 66800935, and 66800936 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)

Reason

Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo

Action

A product bulletin titled "Soft Port Enhancements" dated March 14, 2013, was released for external distribution describing the changes to the Soft Port products, e.g. increasing the Soft Port orifices size and adding a second slit to the release paper covering the port adhesive, and stating the purpose of the modifications to the device. These product bulletin was provided to sales representatives for distribution to customers requesting information regarding the design enhancement made to the Soft Port product range. The bulletin was not distributed to all the customers. To ensure that the product modification and its purpose was appropriately communicated to consignees of affected product a Medical Device Correction Notice, dated 01/22/2014, retrospectively advising customers of the modification to the Soft Port products has been prepared and was sent to consignees of the affected products on the same date.

Distribution

Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, KY, LA, MA, MS, MT, NC, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV, HI and the countries Dubai, Curacao, Argentina, Australia, Chile, Germany, France, England, Hong Kong, Kuwait, Lebanon, Libya, Sri Lanka, Malaysia, New Zealand, Manila, Saudi Arabia, South Africa, Singapore, Turkey, Taipei and Vietnam.

Quantity

498,980 individuals units