BIOPLEX 2200 MULTI-ANALYTE DETECTION SYSTEM: a) 100 test - Catalog No. 665-1150; and b) the 5,000 tests - Catalog No. 665-1155 - manufactured by Bio-Rad Laboratories in Redmond, WA. The product is used as an aid in the diagnosis of systemic autoimmune diseases.
Recall
- Recall Number
- Z-1159-2008
- Event Number
- 45875
- Firm
- Bio-Rad Laboratories
- FEI Number
- 3008272466
- Product Code
- LKJ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- November 13, 2007
- Posted
- June 11, 2008
- Terminated
- May 23, 2011
- Address
- 6565 185th Ave NE, Redmond, WA, 98052-5039
Description
BIOPLEX 2200 MULTI-ANALYTE DETECTION SYSTEM: a) 100 test - Catalog No. 665-1150; and b) the 5,000 tests - Catalog No. 665-1155 - manufactured by Bio-Rad Laboratories in Redmond, WA. The product is used as an aid in the diagnosis of systemic autoimmune diseases.
False negative results due to reagent packs exhibiting low signal.
On 11/13/2007, the firm faxed Voluntary Field Correction Letters (which are dated 11/9/2007) to its consignees and also contacted them via telephone. The Voluntary Field Correction Letters: a) informed them of the potential for false negative results; b) recommended that they perform QC testing daily with each reagent pack; c) provided a new insert for the Instructions for Use; and d) provided a Response Sheet for each consignee to complete and return to the firm.
Worldwide Distribution -- USA, Canada, France, United Kingdom, Italy, Germany, South Africa, and Switzerland.
8,804 kits