FDA Recall Terminated

BIOPLEX 2200 MULTI-ANALYTE DETECTION SYSTEM: a) 100 test - Catalog No. 665-1150; and b) the 5,000 tests - Catalog No. 665-1155 - manufactured by Bio-Rad Laboratories in Redmond, WA. The product is used as an aid in the diagnosis of systemic autoimmune diseases.

Recall: Z-1159-2008 · Initiated November 13, 2007

Recall

Recall Number
Z-1159-2008
Event Number
45875
Firm
Bio-Rad Laboratories
FEI Number
3008272466
Product Code
LKJ
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
November 13, 2007
Posted
June 11, 2008
Terminated
May 23, 2011
Address
6565 185th Ave NE, Redmond, WA, 98052-5039

Description

BIOPLEX 2200 MULTI-ANALYTE DETECTION SYSTEM: a) 100 test - Catalog No. 665-1150; and b) the 5,000 tests - Catalog No. 665-1155 - manufactured by Bio-Rad Laboratories in Redmond, WA. The product is used as an aid in the diagnosis of systemic autoimmune diseases.

Reason

False negative results due to reagent packs exhibiting low signal.

Action

On 11/13/2007, the firm faxed Voluntary Field Correction Letters (which are dated 11/9/2007) to its consignees and also contacted them via telephone. The Voluntary Field Correction Letters: a) informed them of the potential for false negative results; b) recommended that they perform QC testing daily with each reagent pack; c) provided a new insert for the Instructions for Use; and d) provided a Response Sheet for each consignee to complete and return to the firm.

Distribution

Worldwide Distribution -- USA, Canada, France, United Kingdom, Italy, Germany, South Africa, and Switzerland.

Quantity

8,804 kits