FDA Recall
Terminated
Alsius CoolGuard 3000 Temperature Regulation System, Model CoolGuard 3000
Recall: Z-1159-04
·
Initiated June 21, 2004
Recall
- Recall Number
- Z-1159-04
- Event Number
- 29538
- Firm
- Alsius Corporation
- FEI Number
- 3002922896
- Product Code
- NCX
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 21, 2004
- Posted
- July 23, 2004
- Terminated
- December 2, 2004
- Address
- 15770 Laguna Canyon Rd, Ste 150, Irvine, CA, 92618-3111
Description
Alsius CoolGuard 3000 Temperature Regulation System, Model CoolGuard 3000
Reason
Higher than expected rate of pump alarms occur under specific conditions.
Action
Firm replaced circuit board having a new resistor. Recall is complete.
Distribution
MA, PA, VA, FL, AL, MO, TX
Quantity
15