Clearlink System Vented Paclitaxel Set, 10 drops/mL, 111'', Polyethylene Lined Tubing, Non-DEHP Pump Segment, 0.22 Micron Extended Life Filter, Luer Activated Valve, Male Luer Lock Adapter, catalog 2C8857; Baxter Healthcare Corporation, Deerfield, IL 60015 USA
Recall
- Recall Number
- Z-1158-03
- Event Number
- 26772
- Firm
- Baxter Healthcare Corp. Rt
- FEI Number
- 1416980
- Product Code
- FPA
- Status
- Terminated
- Root Cause
- Other
- Initiated
- July 16, 2003
- Posted
- August 21, 2003
- Terminated
- June 29, 2004
- Address
- 120 & Wilson Rd, Round Lake, IL, 60073
Description
Clearlink System Vented Paclitaxel Set, 10 drops/mL, 111'', Polyethylene Lined Tubing, Non-DEHP Pump Segment, 0.22 Micron Extended Life Filter, Luer Activated Valve, Male Luer Lock Adapter, catalog 2C8857; Baxter Healthcare Corporation, Deerfield, IL 60015 USA
There are possible pinholes or perforations in the set tubing, caused by damage that occurred during the packaging process.
Recalled by letter dated 7/16/03. The accounts were informed of the potential for leaks in the set tubing, and asked to inspect their inventory for the affected product codes and lots, segregate any affected products found and discard them according to local procedures. The customers will receive credit for the discarded affected product.
Nationwide, Mexico, Canada, Columbia, Australia, New Zealand and Taiwan
11,280 units