FDA Recall Terminated

Panaroma Central Station Product Usage: This device can view real time, store, print, graph, and trend patient clinical and demographic data. The device can also set independent alarm limits for data send by the bedside monitor.

Recall: Z-1156-2014 · Initiated January 8, 2014

Recall

Recall Number
Z-1156-2014
Event Number
67495
Firm
Mindray DS USA, Inc. dba Mindray North America
FEI Number
2221819
Product Code
MHX
Status
Terminated
Root Cause
Software design
Initiated
January 8, 2014
Posted
March 7, 2014
Terminated
January 21, 2015
Address
800 MacArthur Blvd, Mahwah, NJ, 07430-2001

Description

Panaroma Central Station Product Usage: This device can view real time, store, print, graph, and trend patient clinical and demographic data. The device can also set independent alarm limits for data send by the bedside monitor.

Reason

Mindray has identified two issues with the Panorama Central Station. Full Disclosure data may lag behind the Panorama system time clock when the Panorama Central Station is in use with Mindrays V-Series Monitor. Additionally, event indicators lag behind the Panorama system time clock when the Panorama Central Station is in use with a Panorama Telepack.

Action

Mindray DS USA Inc. sent an Urgent Corrective Action letter dated January 8, 2014 via certified mail to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to contact Mindray Service Representative, at 1-800-288-2121 (Monday Friday 8:30 am-5:30 pm), to arrange for the software update of the Panorama System(s) in their facility. This work will be performed at no cost.

Distribution

Worldwide Distribution - USA Nationwide Distribution

Quantity

732 units in the US