Panaroma Central Station Product Usage: This device can view real time, store, print, graph, and trend patient clinical and demographic data. The device can also set independent alarm limits for data send by the bedside monitor.
Recall
- Recall Number
- Z-1156-2014
- Event Number
- 67495
- Firm
- Mindray DS USA, Inc. dba Mindray North America
- FEI Number
- 2221819
- Product Code
- MHX
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- January 8, 2014
- Posted
- March 7, 2014
- Terminated
- January 21, 2015
- Address
- 800 MacArthur Blvd, Mahwah, NJ, 07430-2001
Description
Panaroma Central Station Product Usage: This device can view real time, store, print, graph, and trend patient clinical and demographic data. The device can also set independent alarm limits for data send by the bedside monitor.
Mindray has identified two issues with the Panorama Central Station. Full Disclosure data may lag behind the Panorama system time clock when the Panorama Central Station is in use with Mindrays V-Series Monitor. Additionally, event indicators lag behind the Panorama system time clock when the Panorama Central Station is in use with a Panorama Telepack.
Mindray DS USA Inc. sent an Urgent Corrective Action letter dated January 8, 2014 via certified mail to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to contact Mindray Service Representative, at 1-800-288-2121 (Monday Friday 8:30 am-5:30 pm), to arrange for the software update of the Panorama System(s) in their facility. This work will be performed at no cost.
Worldwide Distribution - USA Nationwide Distribution
732 units in the US