FDA Recall Open, Classified

The INFX-8000F is designed to take advantage of the latest technological innovations to reduce dose and to save time. A revolutionary graphic user interface and a multi-tasking computer enable the system to fully meet the requirement for optimum image quality, safety, ease of use, advanced efficiency and improved workflow. The advanced floormounted system INFX-8000F uses a new flat panel detector to present fluoroscopic and fluorographic images of high quality.

Recall: Z-1155-2024 · Initiated November 15, 2023

Recall

Recall Number
Z-1155-2024
Event Number
93844
Firm
Canon Medical System, USA, INC.
FEI Number
2020563
Product Code
OWB
Status
Open, Classified
Root Cause
Component change control
Initiated
November 15, 2023
Posted
February 16, 2024
Address
2441 Michelle Dr, Tustin, CA, 92780-7047

Description

The INFX-8000F is designed to take advantage of the latest technological innovations to reduce dose and to save time. A revolutionary graphic user interface and a multi-tasking computer enable the system to fully meet the requirement for optimum image quality, safety, ease of use, advanced efficiency and improved workflow. The advanced floormounted system INFX-8000F uses a new flat panel detector to present fluoroscopic and fluorographic images of high quality.

Reason

VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.

Action

On 11/15/23 correction notices were distributed to customers who were asked to do the following: 1) Should this issue occur please attempt to restart the system and contact your service representative. 2) A recalling firm representative will contact you to schedule a time to replace the flat panel detector on the console. The parts required to address the defect are expected to be available no earlier that December 2023. 3) Complete and return the customer reply form via email to [email protected] For questions, please contact our InTouch team at [email protected] or 800-421-1968, or 800.521.1968.

Distribution

US: CA, LA

Quantity

2