FDA Recall Terminated

RoboCouch Patient Positioning System Product Usage: is intended for use in the support and positioning of a patient during radiosurgery and radiotherapy procedures and other medical procedures when precise positioning is required.

Recall: Z-1155-2013 · Initiated March 27, 2013

Recall

Recall Number
Z-1155-2013
Event Number
64809
Firm
Accuray Incorporated
FEI Number
1000123585
Product Code
JAI
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
March 27, 2013
Posted
April 20, 2013
Terminated
August 20, 2013
Address
1310 Chesapeake Ter, Sunnyvale, CA, 94089-1100

Description

RoboCouch Patient Positioning System Product Usage: is intended for use in the support and positioning of a patient during radiosurgery and radiotherapy procedures and other medical procedures when precise positioning is required.

Reason

A5 gearboxes responsible for pitch (head up and head down) and roll (patient left and right) motions of the RoboCouch manipulator may be defective.

Action

Accuray issued an Urgent Advisory Notification letter dated March 27, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated that Field Service Engineers will be replacing the affected product. Questions and concerns are to be directed to Accuray Customer Support at 1-877-668-8667

Distribution

Worldwide Distribution - USA Nationwide including the states of: OH, MO. OK, CA and VA and the countries of: Finland, France, Luxembourg, Poland, Russia, Turkey and Japan.

Quantity

16 devices are subject to correction.