RoboCouch Patient Positioning System Product Usage: is intended for use in the support and positioning of a patient during radiosurgery and radiotherapy procedures and other medical procedures when precise positioning is required.
Recall
- Recall Number
- Z-1155-2013
- Event Number
- 64809
- Firm
- Accuray Incorporated
- FEI Number
- 1000123585
- Product Code
- JAI
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- March 27, 2013
- Posted
- April 20, 2013
- Terminated
- August 20, 2013
- Address
- 1310 Chesapeake Ter, Sunnyvale, CA, 94089-1100
Description
RoboCouch Patient Positioning System Product Usage: is intended for use in the support and positioning of a patient during radiosurgery and radiotherapy procedures and other medical procedures when precise positioning is required.
A5 gearboxes responsible for pitch (head up and head down) and roll (patient left and right) motions of the RoboCouch manipulator may be defective.
Accuray issued an Urgent Advisory Notification letter dated March 27, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated that Field Service Engineers will be replacing the affected product. Questions and concerns are to be directed to Accuray Customer Support at 1-877-668-8667
Worldwide Distribution - USA Nationwide including the states of: OH, MO. OK, CA and VA and the countries of: Finland, France, Luxembourg, Poland, Russia, Turkey and Japan.
16 devices are subject to correction.