ALM Prismatic Surgical Light 9551 DF; Maquet SA. ALM PRISMATIC (PRC) Surgical Lights are used to provide visible illumination for the surgical field or for the examination of the patient. PRC is designed to eliminate shadow, reduce reflected heat, and illuminate surgical incisions of all depths during surgical procedures.
Recall
- Recall Number
- Z-1153-2009
- Event Number
- 51297
- Firm
- MAQUET Inc.
- FEI Number
- 3008355164
- Product Code
- FTD
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- February 23, 2009
- Posted
- April 12, 2009
- Terminated
- September 29, 2009
- Address
- 1140 US Highway 22 East, Suite 202, Bridgewater, NJ, 08807-2958
Description
ALM Prismatic Surgical Light 9551 DF; Maquet SA. ALM PRISMATIC (PRC) Surgical Lights are used to provide visible illumination for the surgical field or for the examination of the patient. PRC is designed to eliminate shadow, reduce reflected heat, and illuminate surgical incisions of all depths during surgical procedures.
The screws holding the Prismatic triple surgical lights may break due to dynamic multidirectional bending stress. The firm has received one complaint report.
An "Urgent Device Correction" letter dated February 23, 2009 was mailed by FedEx Tracking to all affected consignees. The letter described the issue, potential hazard and recommended precaution. Customers were instructed to complete the enclosed "Customer Response Fax Form" and return it via fax (1-732-667-1903) to Maquet, Inc. Maquet, Inc. will schedule repair service of the affected units once the "Customer Response Fax Form" is received. Direct questions to your local Maquet, Inc. Representative.
Nationwide Distribution.
54 of all units, both article numbers