ALM Prismatic Surgical Light 9551 DF; Maquet SA. ALM PRISMATIC (PRC) Surgical Lights are used to provide visible illumination for the surgical field or for the examination of the patient. PRC is designed to eliminate shadow, reduce reflected heat, and illuminate surgical incisions of all depths during surgical procedures.

Recall: Z-1153-2009 · Initiated February 23, 2009

Recall

Recall Number
Z-1153-2009
Event Number
51297
Firm
MAQUET Inc.
FEI Number
3008355164
Product Code
FTD
Status
Terminated
Root Cause
Device Design
Initiated
February 23, 2009
Posted
April 12, 2009
Terminated
September 29, 2009
Address
1140 US Highway 22 East, Suite 202, Bridgewater, NJ, 08807-2958

Description

ALM Prismatic Surgical Light 9551 DF; Maquet SA. ALM PRISMATIC (PRC) Surgical Lights are used to provide visible illumination for the surgical field or for the examination of the patient. PRC is designed to eliminate shadow, reduce reflected heat, and illuminate surgical incisions of all depths during surgical procedures.

Reason

The screws holding the Prismatic triple surgical lights may break due to dynamic multidirectional bending stress. The firm has received one complaint report.

Action

An "Urgent Device Correction" letter dated February 23, 2009 was mailed by FedEx Tracking to all affected consignees. The letter described the issue, potential hazard and recommended precaution. Customers were instructed to complete the enclosed "Customer Response Fax Form" and return it via fax (1-732-667-1903) to Maquet, Inc. Maquet, Inc. will schedule repair service of the affected units once the "Customer Response Fax Form" is received. Direct questions to your local Maquet, Inc. Representative.

Distribution

Nationwide Distribution.

Quantity

54 of all units, both article numbers