FDA Recall Open, Classified

Epidural tray. PERIFIX 17 Ga. x 3-1/2 in. (90 mm) Tuohy, 5 mL Glass Luer Lock LOR Tray (Kit) A collection of devices intended for the administration of an analgesic or anesthetic agent to the epidural space for the management of pain. It includes the pharmaceutical anesthetic agent and a needle, with additional devices utilized for the procedure (e.g., syringe, dressings), and may include a catheter (i.e., may be intended for single- or continuous-administration). It is not intended for intrathecal anesthesia. Product Code: 332248

Recall: Z-1150-2024 · Initiated January 22, 2024

Recall

Recall Number
Z-1150-2024
Event Number
93889
Firm
B. Braun Medical, Inc.
FEI Number
2523676
Product Code
OGE
Status
Open, Classified
Root Cause
Process control
Initiated
January 22, 2024
Posted
February 16, 2024
Address
901 Marcon Blvd, Allentown, PA, 18109-9512

Description

Epidural tray. PERIFIX 17 Ga. x 3-1/2 in. (90 mm) Tuohy, 5 mL Glass Luer Lock LOR Tray (Kit) A collection of devices intended for the administration of an analgesic or anesthetic agent to the epidural space for the management of pain. It includes the pharmaceutical anesthetic agent and a needle, with additional devices utilized for the procedure (e.g., syringe, dressings), and may include a catheter (i.e., may be intended for single- or continuous-administration). It is not intended for intrathecal anesthesia. Product Code: 332248

Reason

Kits should contain Filter Straws with Standard Luer Connections however, the kits contain Filter Straws with Neuraxial connectors preventing interface with the other components of the kit.

Action

B.Braun issued Urgent Medical Device Recall letter on Monday, January 22, 2024. Letter states reason for recall, health risk and action to take: 1. Review the Device Recall Notification and ensure that all users of the product in your organization, and other concerned personnel are informed about this recall. If you are a distributor and have further distributed the product, please forward this notice to your consignees. The recall is to be extended to the user level. 2. Determine your current inventory of the affected batches within inventory of your facility, cease use and quarantine product subject to recall. Do not destroy any affected product. 3. Record the total number of individual impacted units on the Recall Acknowledgement Form. If you have no inventory remaining, please enter zero (0) on the form. 4. Return the completed Recall Acknowledgement Form to B. Braun Medical Inc. Quality Assurance department by faxing the form to (610) 849-1197 or e-mail to [email protected] within two (2) weeks of receipt, even if the total inventory in your possession is zero (0). 5. Once we receive your Recall Acknowledgement Form, a B. Braun Customer Support representative will contact you with instructions on how to return any impacted cases, including partial cases, in your possession and provide credit and/or replacement of the product based on your individual needs.

Distribution

US Nationwide

Quantity

3,890 units