FDA Recall Open, Classified

Atellica IM Folate 140 test kit- For in vitro diagnostic use in the quantitative determination of folate in human serum and red blood cells. Siemens Material Number (SMN): 10995572

Recall: Z-1149-2023 · Initiated January 9, 2023

Recall

Recall Number
Z-1149-2023
Event Number
91522
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
1219913
Product Code
CGN
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
January 9, 2023
Posted
February 21, 2023
Address
333 Coney St, East Walpole, MA, 02032-1516

Description

Atellica IM Folate 140 test kit- For in vitro diagnostic use in the quantitative determination of folate in human serum and red blood cells. Siemens Material Number (SMN): 10995572

Reason

Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.

Action

Siemens Healthineers issued Urgent Medical Device Corrections (UMDC AIMC 23-03.A-1.US and AIMC 23-03.A-2.US) to US customers and Urgent Field Safety Notices (UFSN AIMC 23-03.A-1.OUS and AIMC 23-03.A-2.OUS) to customers outside the United States. The UMDCs and UFSNs advise customers to use the appropriate FolSerum calibration instructions when testing serum samples with Atellica IM Folate kit lots ending in 337 and above or ADVIA Centaur Folate kit lots ending in 336 and above. Letter states reason for recall, health risk and action to take: Follow the instructions in this Urgent Medical Device Correction until the Atellica IM Folate/ADVIA Centaur Folate Instructions for Use are updated. o Before updating to the new reagent lot, ensure you have processed all serum samples necessary for lot-to-lot comparisons using your existing inventory. o Once you have updated to the new reagent lot, you will not be able to process serum samples with the previous reagent lot. o If your laboratory runs both serum and whole blood sample types, then both assays must be calibrated with the new reagent lot. o If your laboratory runs one sample type, only calibrate the sample type you use If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.

Distribution

Nationwide Foreign: Albania¿¿¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿ Belgium¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Bulgaria¿¿¿¿¿¿¿ Croatia¿¿¿¿¿¿¿¿ Cura¿ao,¿St¿Eus Cyprus¿¿¿¿¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ Estonia¿¿¿¿¿¿¿¿ Finland¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Hungary¿¿¿¿¿¿¿¿ Iran¿¿¿¿¿¿¿¿¿¿¿ Ireland¿¿¿¿¿¿¿¿ Israel¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Jordan¿¿¿¿¿¿¿¿¿ Kenya¿¿¿¿¿¿¿¿¿¿ Latvia¿¿¿¿¿¿¿¿¿ Lithuania¿¿¿¿¿¿ Malta¿¿¿¿¿¿¿¿¿¿ Morocco¿¿¿¿¿¿¿¿ Nepal¿¿¿¿¿¿¿¿¿¿ Netherlands¿¿¿¿ Norway¿¿¿¿¿¿¿¿¿ Oman¿¿¿¿¿¿¿¿¿¿¿ Pakistan¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Qatar¿¿¿¿¿¿¿¿¿¿ Romania¿¿¿¿¿¿¿¿ Saudi¿Arabia¿¿¿ Serbia¿¿¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Slovenia¿¿¿¿¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Sweden¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Tunisia¿¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ U.A.E.¿¿¿¿¿¿¿¿¿ Uganda¿¿¿¿¿¿¿¿¿ Ukraine¿¿¿¿¿¿¿¿ United¿Kingdom¿ Vatikancity¿¿¿¿ Country AR AU BR BS CA CL CN CO CR DO EC HK IL IN JP KR MX MY NZ PE PH PY SA SG TH TW UY VN ZA

Quantity

13,843 kits