Product 49 consists of all product under product code: JWH and same usage: Item no: 597206529 NEXGEN ALL-POLY PATELLA, 597206532 NEXGEN ALL-POLY PATELLA, 597206535 NEXGEN ALL-POLY PATELLA, 597206538 NEXGEN ALL-POLY PATELLA, 597206541 NEXGEN ALL-POLY PATELLA, 598801713 NEXGEN CEMENTED STEM EXT 598801813 NEXGEN CEMENTED STEM EXT 598801913 NEXGEN CEMENTED STEM EXT 598801610 NEXGEN FLUTED STEM EXT, 1 598801611 NEXGEN FLUTED STEM EXT, 1 598801612 NEXGEN FLUTED STEM EXT, 1 598801613 NEXGEN FLUTED STEM EXT, 1 598801614 NEXGEN FLUTED STEM EXT, 1 598801615 NEXGEN FLUTED STEM EXT, 1 598801616 NEXGEN FLUTED STEM EXT, 1 598801618 NEXGEN FLUTED STEM EXT, 1 598801619 NEXGEN FLUTED STEM EXT, 1 598801620 NEXGEN FLUTED STEM EXT, 2 598801622 NEXGEN FLUTED STEM EXT, 2 598801624 NEXGEN FLUTED STEM EXT, 2 597206126 NEXGEN MICRO ALL-POLY PAT 597206129 NEXGEN MICRO ALL-POLY PAT 597206132 NEXGEN MICRO ALL-POLY PAT 597206135 NEXGEN MICRO ALL-POLY PAT for use in total knee arthroplasty
Recall
- Recall Number
- Z-1147-2016
- Event Number
- 72900
- FEI Number
- 1000132037
- Product Code
- JWH
- Status
- Terminated
- Root Cause
- Packaging
- Initiated
- January 11, 2016
- Posted
- March 12, 2016
- Terminated
- April 13, 2017
- Address
- Zimmer Manufacturing B.V. Turpeaux Industrial Park Rd 1 km 123.4, Mercedita, PR, 00715
Description
Product 49 consists of all product under product code: JWH and same usage: Item no: 597206529 NEXGEN ALL-POLY PATELLA, 597206532 NEXGEN ALL-POLY PATELLA, 597206535 NEXGEN ALL-POLY PATELLA, 597206538 NEXGEN ALL-POLY PATELLA, 597206541 NEXGEN ALL-POLY PATELLA, 598801713 NEXGEN CEMENTED STEM EXT 598801813 NEXGEN CEMENTED STEM EXT 598801913 NEXGEN CEMENTED STEM EXT 598801610 NEXGEN FLUTED STEM EXT, 1 598801611 NEXGEN FLUTED STEM EXT, 1 598801612 NEXGEN FLUTED STEM EXT, 1 598801613 NEXGEN FLUTED STEM EXT, 1 598801614 NEXGEN FLUTED STEM EXT, 1 598801615 NEXGEN FLUTED STEM EXT, 1 598801616 NEXGEN FLUTED STEM EXT, 1 598801618 NEXGEN FLUTED STEM EXT, 1 598801619 NEXGEN FLUTED STEM EXT, 1 598801620 NEXGEN FLUTED STEM EXT, 2 598801622 NEXGEN FLUTED STEM EXT, 2 598801624 NEXGEN FLUTED STEM EXT, 2 597206126 NEXGEN MICRO ALL-POLY PAT 597206129 NEXGEN MICRO ALL-POLY PAT 597206132 NEXGEN MICRO ALL-POLY PAT 597206135 NEXGEN MICRO ALL-POLY PAT for use in total knee arthroplasty
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to [email protected]. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.
US Nationwide Distribution
53338